Transcription
Long-Term Shutdown Quick Reference GuideOverviewThis guide contains instructions for performing a long-term shutdown for the following BD flow cytometers. BD Accuri C6, C6 Plus and BD FACSVia Cytometers BD FACS Lyse Wash Assistant Systems BD FACSAria Family of Cell Sorters BD FACSMelody Cell Sorters BD FACSCalibur Cytometers BD FACS Sample Prep Assistant Systems BD FACSCanto Family of Cytometers BD Influx and BD FACSJazz Cell Sorters BD FACSLyric and BD FACSVerse Cytometers BD LSRFortessa , BD FACSCelesta and BD FACSymphony Family of CytometersGeneral RecommendationsPerform these procedures when shutting down your instrument for long term-storage ( one week or more, unlessotherwise specified in the Instructions for Use or User Guide).When shutting down an instrument for long term-storage, perform standard cleaning or decontaminationprocedures and leave the instrument in deionized (DI) water or the recommended storage solution.For any system, we recommend: Use absolute ethanol, freshly diluted down to 70%. Do not use denatured ethanol or any methanol-containing fluids inyour cytometer. Use freshly diluted 10% bleach for maximum efficacy. Use sterile deionized water, not distilled. Where noted, use filters that are dedicated for the specific fluid. Dispose of all liquid waste according to your standard biohazard protocols.BD Accuri C6, C6 Plus, and BD FACSVia Cytometers1Perform a SIP clean using 2 mL of BD FACS Clean Solution, followed by 2 mL of DI water.2Remove the fluid bottles and rinse thoroughly with DI water.3Refill all 3 bottles with DI water.4Start up the cytometer and then shut it down (2 cycles).For more information, see the Maintenance chapter in the BD Accuri C6 Plus System User Guide.Procedure: SIP Clean, Starting up the system, Shutting down the system.For more information, see the Maintenance chapter in the BD FACSVia Instructions for Use.Procedure: SIP Clean, Starting up the system, Shutting down the system.BD Accuri C6 is discontinued.BD Accuri C6 and BD Accuri C6 Plus are for Research Use Only. Not for use in diagnostic or therapeutic procedures.BD FACSVia System is CE marked in compliance with the European In Vitro Diagnostic Medical Device Directive 98/79/EC.1
BD FACSAria II, III, Fusion, and BD FACSymphony S6 Cell Sorters1Refill each 5-L tank with the appropriate cleaning solution, using fresh 10% bleach, 70% ethanol, andsterile DI water.2Perform a Prepare for Aseptic Sort cleaning mode. Select Cytometer Cleaning Modes Prepare for Aseptic Sort and follow the on-screen prompts to thoroughlyclean all lines. When the procedure is complete, install a tube of sterile, filtered DI water on the loading port and clean the flow cell.3Replace bleach in the tank and lines with DI water. Empty the 5-L bleach tank and refill with sterile DI water. Connect the tank and select Cytometer Prime After Tank Refill and select Bleach to clear bleach from the lines.For more information, see the Maintenance chapter in the BD FACSAria II, III, BD FACSAria Fusion. or BDFACSymphony S6 User Guide.Procedure: Prepare for Aseptic Sort.BD FACSAria II is discontinued.BD FACSAria II, BD FACSAria III, BD FACSAria Fusion and BD FACSymphony S6 are for Research Use Only.Not for use in diagnostic or therapeutic procedures.BD FACSCalibur Systems1Run a monthly cleaning procedure. Fill a spare reservoir with 1-2 liters of 10% bleach and install in the fluidics drawer in place of the sheath fluid reservoir. Bypass the sheath filter. NOTE: Do not run bleach or detergent through the sheath filter. Disconnect the top of the sheath filter tubing from the filter port and replace it with the sheath tubing from thespare reservoir. Install and run a tube containing 3 mL of 10% bleach on the SIP and run on HI for 20-30 minutes. Replace the bleach reservoir with a reservoir containing deionized (DI water) and run a tube of DI water for 20-30 minutes.2Leave the filter bypass in place and DI water in the fluidics system until you are ready to use the cytometer again.For more information, see the Maintenance chapter in the BD FACSCalibur Instructions for Use (IFU).Procedure: Decontaminating the Fluidics.BD HTS option3If you have the BD HTS option, flush the HTS with DI water. Loosen the positioning screw underneath the support bracket and pull the HTS slightly forward until you can access theconnections on the back of the unit. Disconnect the sheath line on the back of the HTS. Connect the purge line to the sheath connection on the HTS and place the other end in a beaker of DI water. Place the cover back on the HTS and select HTS Prime. Repeat the HTS prime 9 times to completely fill the HTS with DI water.Leave the HTS with DI water in the lines and turn off the system.For more information, see the Maintenance chapter in the BD High Throughput Sampler User Guide.Procedure: Placing the HTS in Long-Term Storage.2
BD FACSFlow Supply System4If you have the BD FACSFlow supply system, rinse the tanks and lines with DI water. Empty and rinse the plenum and fill with DI water. Remove the probe from the 20-L sheath container and place in a tall graduated cylinder full of DI water. Press and hold the prime button on the BD FACSFlow supply system for 20 seconds to flush the lines with DI water. Empty the plenum and let it dry.For more information, see the System Overview chapter in the BD FACSFlow Supply System User Guide for adescription of each component.BD FACSCalibur instruments is discontinued. At the time the BD FACSCalibur instrument was placed on the Union market, theinstrument was CE marked in compliance with the European In Vitro Diagnostic Medical Device Directive 98/79/EC.BD High Throughput Sampler and BD FACSFlow Supply System are for Research Use Only.Not for use in diagnostic or therapeutic procedures.BD FACSCanto II and 10-color Configuration Systems1If you have the BD HTS option, connect the HTS sample coupler to the SIT and close the HTS door.2Refill fluids and empty waste, if needed, and then select Cytometer Long Clean.3When the long clean is complete, run a fluidics shutdown and turn off the system.4Wipe down the SIT and aspirator arm with DI water.For more information, see the Maintenance chapter in the BD FACSCanto II or BD FACSCanto 10-color Instructions for Use.Procedure: Decontaminating the Fluidics System for Storage.BD FACSCanto II Flow Cytometer is CE marked in compliance with the European In Vitro Diagnostic Medical Device Directive98/79/EC.BD FACSCanto Flow Cytometer (10-color configuration) is discontinued in EU. At the time the BD FACSCanto Flow Cytometer(10-color configuration) was placed on the Union market, the instrument was CE marked in compliance with the European InVitro Diagnostic Medical Device Directive 98/79/EC.BD FACSLyric and BD FACSVerse Systems1Select Cytometer Monthly Clean to start the monthly clean procedure. Follow the instructions in the Monthly Clean dialog.2Stop when you reach the step that instructs you to refill the sheath tank with sheath fluid, and refill the sheath tankwith DI water or BD FACS Shutdown Solution.3Complete the procedure in the Monthly Clean dialog, but leave the sheath filter bypass line in place.4Wipe down the exterior surfaces with a 10% bleach solution, then wipe them down again with DI water.For more information, see the Maintenance chapter in the BD FACSLyric Instructions for Use or BD FACSVerseReference System User Guide.Procedure: Preparing for long-term shutdown.BD FACSLyric Flow Cytometer is CE marked in compliance with the European In Vitro Diagnostic Medical Device Directive98/79/EC.BD FACSVerse is discontinued.BD FACSVerse is for Research Use Only. Not for use in diagnostic or therapeutic procedures.3
BD FACS Lyse Wash Assistant12Empty the waste tank of any ammonium chloride waste and rinse with water.Run a Wash Only protocol. Install 3 tubes of 10% bleach into positions 1-3 on the carousel. Install 3 tubes of DI water into positions 4-6 on the carousel. Select Wash Only.3Empty and rinse all tanks and fill with DI water.4Run a Duo-Lyse protocol with empty tubes to flush the system with DI water. Install 10 empty tubes into positions 1-10 on the carousel. Select Duo-Lyse. Remove rack after processing is complete.5Shut down the instrument and empty, rinse and air dry all the tanks. Wipe down all exposed surfaces with 10%bleach followed by DI water.For more information, see the Maintenance chapter in the BD FACS Lyse Wash Assistant User Guide.Procedure: Daily Maintenance and Shutdown.BD FACS Lyse Wash Assistant is CE marked in compliance with the European In Vitro Diagnostic Medical Device Directive98/79/EC.BD FACSMelody Cell Sorters1Add at least 2.5 L of 70% ethanol to the sheath tank (orextra tank if purchased).NOTE: Do not use tanks from a different system. Onlyuse tanks provided with the BD FACSMelody System.2On the Cytometer page of the software, selectLong-Term Shutdown and complete the stepsin the wizard.3Power off the system and computer.4Perform an extended fluidics startup when youresume using the cell sorter.For more information, see the System startup and shutdown chapter in the BD FACSMelody User Guide.Procedure: Shutting down the system, Run Extended Fluidics Startup.BD FACSMelody is for Research Use Only. Not for use in diagnostic or therapeutic procedures.4
BD FACS Sample Prep Assistant (SPA) II and III Systems1Perform a weekly cleaning procedure. Select Instrument Clean Weekly Clean. Place a vial containing 60 mL of BD FACSClean Solution or 10% bleach in well R25 on the reagent rack. When prompted, clean the Z-rack with 70% ethanol. Select Instrument Home to return the probe to the wash station.2Clean the waste in-line filter assembly.3Prime the instrument with DI water. Empty all tanks except for the waste tank and fill with DI water. Select Instrument Prime.4Shut down the instrument.5Empty and rinse all tanks except the BD FACSFlow Cubitainer and wipe down all exposed surfaces with 10% bleachfollowed by DI water.For more information, see the Shutdown and Maintenance chapter in the BD FACS Sample Prep Assistant III User Guide.Procedures: Weekly Cleaning and Monthly Servicing of Waste In-Line Filters.BD FACS Sample Prep Assistant III is CE marked in compliance with the European In Vitro Diagnostic Medical Device Directive98/79/EC.BD FACS Sample Prep Assistant II is discontinued. At the time BD FACS Sample Prep Assistant II was placed on the Unionmarket, the instrument was CE marked in compliance with the European In Vitro Diagnostic Medical Device Directive 98/79/EC.BD Influx and BD FACSJazz Systems1Clean the sample line. Run a tube of 10% bleach for 5 minutes. Run a tube of DI water for 5 minutes.2Rinse the system with DI water. Fill the debubble reservoir with DI water and install it under the nozzle tip. Fill the sheath tank with 1 L of DI water. Press Rinse and Backflush to rinse all fluid lines with DI water until the tank runs dry.3Dry the fluidics lines. Remove the nozzle tip and place the flush bucket under the nozzle. Disconnect the sheath and air lines from the sheath tank and connect them to each other. Press Rinse and Backflush to run air through the system for 10-15 minutes.4Clean the sort chamber with DI water and turn off the system.For more information, see the System Workflow chapter in the BD Influx or the BD FACSJazz User Guide.Procedure: System Shutdown Workflow.BD FACSJazz is discontinued.BD Influx is discontinued in EU.BD Influx and BD FACSJazz are for Research Use Only. Not for use in diagnostic or therapeutic procedures.5
BD LSRFortessa, BD LSRFortessa X-20, BD FACSCelesta and BD FACSymphony Systems1Run a system flush. Bypass the sheath filter. Run BD FACSClean Solution or 10% bleach through the sheath and sample pathway on high for 30 minutes. Run BD FACSRinse Solution through the sheath and sample pathway on high for 30 minutes. Run DI water through the sheath and sample pathway on high for 30 minutes.2Leave the system with DI water in the lines and a tube of DI water on the SIP.For more information, see the Maintenance chapter in the BD LSRFortessa, BD LSRFortessa X-20, BD FACSCelesta,or the BD FACSymphony User Guide.Procedure: Flushing the System.BD HTS option3If you have the BD HTS option, flush the HTS with DI water. Disconnect the sheath line on the back of the HTS. Connect the purge line to the sheath connection on the HTS and place the other end in a beaker of DI water. Place the cover back on the HTS and select HTS Prime. Repeat the HTS prime 9 times to completely fill the HTS with DI water.4Leave the HTS with DI water in the lines and turn off the system.For more information, see the Maintenance chapter in the BD High Throughput Sampler User Guide.Procedure: Placing the HTS in Long-Term Storage.BD FACSFlow Supply System5If you have the BD FACSFlow Supply System, rinse the tanks and lines with DI water. Empty and rinse the plenum and fill with DI water. Remove the probe from the 20-L sheath container and place in a tall graduated cylinder full of DI water. Press and hold the prime button on the BD FACSFlow supply system for 20 seconds to flush the lines with DI water. Empty the plenum and let it dry.For more information, see the System Overview chapter in the BD FACSFlow Supply System User Guide for adescription of each component.BD LSRFortessa , BD LSRFortessa X-20, BD FACSCelesta , BD FACSymphony , BD High Throughput Sampler and BDFACSFlow Supply System are for Research Use Only. Not for use in diagnostic or therapeutic procedures.23-22571-00BD Life Sciences, San Jose, CA 95131, USAbdbiosciences.com/euBD, the BD Logo, BD Accuri, BD FACS, BD FACSAria, BD FACSCalibur, BD FACSCanto, BD FACSCelesta, BD FACSFlow, BDFACSJazz, BD FACSLyric, BD LSRFortessa, BD FACSMelody, BD FACSymphony, BD FACSVerse, BD FACSVia and BD Influx aretrademarks of Becton, Dickinson and Company or its affiliates. 2020 BD. All rights reserved.
At the time the BD FACSCanto Flow Cytometer (10-color configuration) was placed on the Union market, the instrument was CE marked in compliance with the European In Vitro Diagnostic Medical Device Directive 98/79/EC. BD FACSLyric Flow Cytometer is CE marked in compliance with the Europe