PowerPac BasicPower SupplyInstruction ManualFor technical support, call your local Bio-Rad office, or in the U.S., call 1-800-424-6723.

NoticeNo part of this publication may be reproduced or transmitted in any form or by any means,electronic or mechanical, including photocopy, recording, or any information storage orretrieval system, without permission in writing from Bio-Rad.Bio-Rad reserves the right to modify its products and services at any time. This guide issubject to change without notice. Although prepared to ensure accuracy, Bio-Rad assumesno liability for errors or omissions, or for any damage resulting from the application or use ofthis information.Bio-Rad is a trademark of Bio-Rad Laboratories, Inc. in certain jurisdictions.All trademarks used herein are the property of their respective owner.

Table of ContentsPageSafety.iiiSection 11.11.2Introduction .1Overview .1Unpacking .2Section 2Control Features .3Section 3Setup and Operation.5Section 44.14.24.34.4Maintenance and lacing a Fuse .11Firmware Version Number .11Appendix ASpecifications .13Appendix BWarranty and Ordering Information .15

SafetyCaution/Warning!!PowerPac power supplies use high output voltages that are electrically isolated from earthground to minimize the risk of electrical shock to the user. The following guidelines shouldbe observed and followed when using a PowerPac power supply.PowerPac power supplies have passed test for operation at temperatures between 0º and40º C, with relative humidity between 10 and 95% non-condensing. Operating the powersupply outside these conditions is not recommended by Bio-Rad and will void thewarranty.1. To ensure adequate cooling of the power supply, be sure that there is at least 6 cmclearance around the power supply. Do not block the fan vents at the rear of the unit.2. Always connect the power supply to a 3-prong, grounded AC outlet, using the3-prong AC power cord provided with the power supply.3. Bio-Rad electrophoresis cells have molded two-prong plugs that are inserted into thepower supply's high voltage output jacks. These plugs have been IEC 61010-1* certifiedfor safety compliance for use with PowerPac power supplies. Use of other plugs orbanana jacks is done at the user's own risk and is not recommended by Bio-Rad.When inserting and removing the molded two-prong plug, always grasp the plug by themolded support at the rear of the plug. Do not grasp the individual prong ends.4. Do not operate the power supply in extreme humidity ( 95%) or where condensationcan short the internal electrical circuits of the power supply.5. When taking the power supply into a cold room, the unit can be operatedimmediately. However, when removing the power supply from the cold room, let theunit equilibrate to room temperature for a minimum of 2 hours before using it.6. Never connect a high voltage output lead to earth ground. This defeats the floatingelectrical isolation of the power supply and exposes the user to potentially lethal highvoltages.ImportantThis instrument is intended for laboratory use only.This product conforms to the class A standards for Electromagnetic Emissions, intendedfor laboratory equipment applications. It is possible that emissions from this product mayinterfere with some sensitive appliances when placed nearby or on the same circuit asthose appliances. The user should be aware of this potential and take appropriatemeasures to avoid interference.Bio-Rad's PowerPac power supplies are designed and certified to meet IEC 61010-1* safetystandards. Certified products are safe to use when operated in accordance with theinstruction manual. This safety certification does not extend to electrophoresis cells oraccessories that are not IEC 61010-1 certified, even when connected to this power supply.This instrument should not be modified or altered in any way. Alteration of this instrumentwill void the manufacturer's warranty, void the IEC 61010-1 certification, and create apotential safety hazard for the user.Bio-Rad is not responsible for any injury or damage caused by the use of this instrumentfor purposes other than those for which it is intended, or by modifications of theinstrument not performed by Bio-Rad or an authorized agent.*IEC 61010-1 is an internationally accepted electrical safety standard for laboratory instruments.

Section 1Introduction1.1 OverviewThe PowerPac Basic provides constant voltage or constant current to instruments used inelectrophoresis. The power supply operates at the values specified for the constantparameter. However, to prevent damage to the electrophoresis cell, the PowerPac Basicprovides automatic crossover to constant current or constant voltage, depending onwhich set value is first reached. When the set limit of the non-constant parameter isreached, and the power capability of the unit is not exceeded, the power supply willswitch, making the non-constant parameter the new constant parameter.Output specifications:Voltage:Adjustable from 10 to 300 Volts, in 1 volt increments.Current:Adjustable from 4 to 400 milliamperes (mA) in 1 mA increments.Power:75 watts (maximum).Four output jacks:Up to four identical electrophoresis cells can be connected in parallelto the power supply.{FRONT PANELOUTPUT JACKSFig. 1. Front View.AC INLETPOWER SWITCHFUSEFig. 2. Rear View.1

The PowerPac Basic has the following features: Programmable constant voltage or constant current with automatic crossover Timer control from 0 to 999 minutes 3-digit LED display Pause mode for editing running parameters Automatic detection of no load conditions and rapid changes in resistance Power Failure Detection in timed modes allowing completion of run Stackable case with adjustable viewing angle via flip down legs (see Figure 3)LEGS(LOWERED POSITION)Fig. 3. Front View with Legs in Lowered Position.1.2 UnpackingWhen you receive the power supply, carefully inspect the container for any damage whichmay have occurred in shipping. Severe damage to the container may indicate damage tothe power supply itself. If you suspect damage to the unit, immediately file a claim with thecarrier in accordance with their instructions before contacting Bio-Rad Laboratories.After unpacking the PowerPac Basic, remove the plastic film from the translucent greentop case. The plastic film may leave a residue. If so, clean with a soft, damp cloth.Contents include: PowerPac Basic power supply Power cord Instruction manual Warranty card Declaration of conformityIf any part is missing or damaged, contact Bio-Rad Laboratories immediately.2

Section 2Control Features3 DIGIT LEDDISPLAY MODE KEYSCROLL KEYSSTOP KEYCONSTANTPARAMETER KEYRUN AND PAUSE KEYFig. 3. Front Panel.KeyDescriptionConstant Parameter Key: Selects either constant voltage or current. The LED indicates the selected parameter. During a run, maximumpower is indicated when both LED’s are lit.Display Mode Key: Selects the parameter to be displayed (volts, milliamperes or minutes).The LED displays the value of the indicated parameter.Scroll Keys: Changes the value of the selected parameter. If the Scroll Key ispressed constantly for more than 5 units in either direction, /-, thevalues will increase/decrease in increments of 10 to reach the desiredvalue faster.Run & Pause Key: Starts and pauses a run. Pausing allows editing of the constantparameter and the parameter values. Corresponding LED indicates the status of the power supply.Stop key: Stops run. Constant parameter and limit parameter values arepreserved. Timer is reset to zero.3

Section 3Setup and OperationSTEPPROCEDUREDESCRIPTION1.Turn power on.Press the power switch located on the rightside of the unit to the on position.The default setting is constant V, and theLED display shows zero value.To display firmware version number, holddown constant parameter key while turningthe power switch to the on position.2.Connect the electrophoresiscell(s) to the power supply.The power leads are color coordinated tothe output terminals in red and black.indicates high voltages.Note: Power leads must be insertedperpendicular to the curve of the case.Fig. 5. Power leads connected correctly.Fig. 6. Power leads connected incorrectly.3.Press the Constant key to select theconstant parameter, either voltage orcurrent The LED on the parameter keycorresponding to the selected constantparameter will light up. The display showsthe zero value for the constant parameter.Select the constant parameteror4.Enter a value for the constantparameter using the scroll key.Use the scroll key to enter the desired value.Voltage: 10–300V adjustable in 1V increments.Current: 4–400 mA adjustable in 1 mAincrements.Note: If the scroll key is pressed constantlyfor more than 5 units in either direction, /-,the values will increase/decrease inincrements of 10 to reach the desired valuefaster.5.Enter value for limit parameter.For voltsFor milliamperes5Use the parameter key to select the limitparameter. The maximum default value,300V or 400mA, is displayed. Select a suitablelimit value to avoid excessive powerconditions for the application.

STEPPROCEDUREDESCRIPTION6.Programming a timed runUse the parameter key to select time. Thedisplay will show a zero value. Use the scrollkey to enter the desired time up to 999 min.If no time is entered the run will continueuntil the run is terminated by pressing thestop key. When 999 minutes is reached, theclock resets to zero and the run continues.Note: In a timed run the displayed time is theremaining time. Pressing the pause keywill keep the remaining time. Pressingthe stop key will reset the time to zero. 7.In an untimed run the displayed time is theelapsed time. Pressing the pause key willretain the elapsed time.Activating the Power failure detection modeis possible for timed runs only.Optional; available only whena timed run is programmed.Power Failure detection. 8.Make sure the parameter key is in timemode (time LED is lit)Enter the desired time if not previouslyenteredHold down the stop key for 2 secondsuntil the display shows Pfd. This indicatesthe Power Failure Detection is activated. Caution: Always use the stop keyto terminate a run in progress. Useof the power switch to terminate arun in progress is treated as apower failure and the appropriate Note: After completion or termination of aerror code is displayed when the run, the Power Failure Detect mode isunit is turned back onautomatically de-activated. See Section 4.2,Troubleshooting, for details on Power FailureDetect error messages.Certain applications exhibit fluctuations inOptionalDe-activation of change inresistance that can trigger the change inresistance feature.resistance error codes. If this is the case,the change in resistance feature can bede-activated to allow un-interruptedcompletion of a run. Make sure the parameter key is incurrent mode (current LED is lit) Hold down the stop key for 2 sec untilthe display shows dE9Note: After completion or termination of arun the Change in Resistance Detection isautomatically activated.Caution: De-activating this safety featureincreases the chance of electrical hazard.6

STEPPROCEDUREDESCRIPTION9.Start the runPress the run/pause key to start the run.The run LED is lit.10.Viewing and editing optionsduring a run. Viewing: Press the parameter key toview the corresponding value on thedisplayEditing: Edit the constant parametervalue and the time value for timed runsNote:Editing the limit value is possible in thepause mode. To change from an un-timedto a timed run, stop the run and re-program.11.Pause modePress run/pause key during a run to enterthe pause mode.When the pause LED is lit it is possible to: Safely make adjustments to theinstrument connected to the powersupply Edit the values for all parameters Change the constant parameterNote:To change from an un-timed to a timed run,stop the run and re-program.12.End of run.When a run is completed, i.e., a timed runhas ended or an untimed run is stopped, theconstant parameter value and limitparameter value are preserved. The timer isreset to zero. Neither the run nor the pauseLED is lit,indicating that no power is supplied to theoutput jacks.13.Terminating a timed run inprogressPress the stop key to terminate a timed run.The constant parameter value and limitparameter value are preserved. The timer isreset to zero. Neither the run nor the pauseLED is lit indicating that no power is suppliedto the output jacks.14.Powering downPress the stop key before turning the powerswitch to the off position. If this is not done,a power failure will be detected causing anautomatic restart if the Power FailureDetection is enabled.7

Section 4Maintenance and Troubleshooting4.1 MaintenanceThe PowerPac Basic requires little maintenance to assure reliable operation. To clean thecase, first unplug the power supply. Use a damp cloth to wipe down the outer case.4.2 TroubleshootingProblemNo display/lights/fanCause1. No AC power.2. Blown fuse.3. Power switchexercised rapidly tothe on and offpositions.Repeated blown fusesLeads from cell arenot long enough to fitoutput jacksHardware failureOutput terminals for thePowerPac Basic arerecessed 16 mm tomeet safety regulations.Some leads are notlong enough to makeelectrical connection.No load detectedE1 error codedisplayedNote: Use of the PowerPac Adaptorvoids IEC 61010-1 safety provisions.Verify all electrical connections.Verify buffer levels where appropriate. E2 error codedisplayedE3 error codedisplayedE5 error codedisplayedE6 error codedisplayedE7 error codedisplayedInstrument notconnected to PS The current load isbelow 4mAOver current (loadcurrent greater then400 mA)Solution1. Check if PowerPac Basic isunplugged, or problem with ACpower source, or power switch is inoff position.2. Replace fuse. See section 4.3 fordetails.3. The unit needs to be reset. Turnpower switch to the off position,wait 5–10 seconds, then turnpower switch to the on position toresumenormal operation.Contact Bio-Rad Technical Resources.Use the PowerPac Adaptor, whichaccommodates most standard 4 mmbanana plugs, to make a secureelectrical connection.Check for and correct any short circuitor excessive load problem. Excessiveload due to high buffer concentrationwill require the buffer be remade.Then, Presskey to resume the run or, To clear the error code, press anykey (other thenkey).Over voltage (loadTurn power supply off, then on to.voltage over 300 V)reset If problems persists, contactBio-Rad Technical Resources.A power failure occurred during a timed run with Power Fail Detection(PFd) activated, and run is completed. Power Fail detection (Pfd) isde-activated after completion or termination of each run.A power failure occurred during a timed run without Power FailDetection (PFd) activated, and run is not completed.Power Failure occurred during an untimed run or the power switchwas turned off before pressing stop,. and run is not completed.Power Fail detect (Pfd) cannot be activated for untimed runs.9

ProblemE8 error codedisplayedCauseRegulation errorE9 error codedisplayedChange in LoadResistance.The PowerPacE10 error codedisplayedE14 error codedisplayedE15 error codedisplayedE17 error codeE18 error codeE19 error codeE20 error codeE21 error codeE22 error codeE23 error codeE24 error codeE25 error codeSolutionTurn power supply off, then on to resetIf problem persists, contact Bio-RadTechnical ResourcesCheck and correct any potentialresistance problem then, press run/pause key to resume runor, press any key, other than the run/pausekey, to clear the error code.Basic detects drasticchanges in resistancewhich may indicatefailure of the cell’s powerleads or a loose outputconnection.Note: Certain applicationsexhibit fluctuations inresistance that cantrigger the change inresistance error code.If this is the case, thechange in resistancefeature can be deactivatedto allow uninterruptedcompletion of a run.(see Section 3, Step 8)Caution: Deactivating thissafety feature increases thepossibility that a failure of thepower leads or loose outputconnection will not bedetected.Unacceptable value(s)Clear the code by pressing any keyenteredother than the run/pause key.Then, enter values within range ofPowerPac Basic and press therun/pause keyInternal Over VoltagePossible power supply malfunction.Check for and correct problems suchInternal Short Circuitas dirty contacts, frayed wires,excessive buffer concentration. Thenpress any key other than the run/pausekey to clear the code.Hardware failureContact Bio-Rad Technical ResourcesHardware failureHardware failureInstrument is overheating Check fan and vents for obstruction.Regulation errorPower fail while runningExceeded voltage limitsettingExceeded current limitsettingExceeded wattage limitsetting10Turn power supply off, then on to resetPress the stop key, then cycle powerPossible power supply malfunction.Check for and correct problems, suchas dirty contacts, frayed wires, excessivebuffer concentration. Then press stopto clear the error.

4.3 Replacing a FuseIf there is no display, lights, or fan, and the PowerPac Basic is plugged into a working ACpower outlet with the power switch in the on position, the fuse may need to be replaced.1. Disconnect the power cord from the electrical outlet.2. Insert a flat blade screwdriver into the notches of the power entry module's fuseholder to release it. See Figure 5.3. Remove the fuse from the fuse holder. Replace with 2.5A, 250V, 5 X 20 mm,fast-blow fuses (Bio-Rad catalog number 12007433). 14. Re-insert the fuse holder into its position. Press the fuse holder gently until it snaps intoplace.The unit is now ready for use.FUSE DRAWERNOTCHESFig. 5. Rear View Showing Fuse Drawer with Notches.Note: Repeated blowing of the fuse indicates a hardware failure and Bio-Rad should becontacted for repair.4.4 Firmware Version NumberTo display the PowerPac Basic’s firmware version number while the power supply is off,hold down the constant parameter key and concurrently turn the power switch to the onposition. All of the LEDs and segments in the 3 digit LED display will light. Release theconstant parameter key and the firmware version number will then appear for a fewseconds. The power supply is now ready for operation.1Fuse specifications are applicable to products with a serial number of 041BR300000 and higher.11

Appendix ASpecificationsInput PowernominalactualFusesInput Power CordOutput (Programmable)VoltageVoltage AccuracyCurrentCurrent AccuracyPower (maximum)TerminalsTimer ControlRippleLine RegulationLoad RegulationDriftNoiseReadout StabilityVoltsCurrentSafety FeaturesNo load detectionSudden load change detectionOverload/short circuit protectionInput line protectionAuto power up after power failureSafety ComplianceEMIDisplay FunctionsFunction ModesEnvironmentalOperating TemperatureHumidityDimensionsUnit is stackableWeight2100–240 VAC, 50/60 Hz85–264 VAC, 50/60 Hz2.5 A, 250 VAC, 5mm x 20mm, fast-blow3-wire; grounded210 V to 300 V, fully adjustable in 1 V steps 2% of reading or 3 volts, whichever is more4 mA to 400 mA, fully adjustable in 1 mA steps 2% of reading or 3 mA, whichever is more75 W4-pair recessed banana jacks, floating in parallel001 to 999 minutes, fully adjustable 1% @ 300 V and 70 W 1% @ 300 V and 70 W, 85–264 VAC 1% @ 300 V for a 50% change in output load 1% @ after 15 min warmup at 300 V and 70 W55 dBA at 1 meter 1 V 1 mAIndicated by error message on displayIndicated by error message on displayAutomatic power limitFuse on both hot and neutralUser-selectable, setup values maintainedIEC 61010-1Conforms to CE standards for Emissions andImmunity class A, tested only at 230V. SeeDeclaration of Conformity for details. TUV EMCcertification3-digit LED displays voltage, current, or timeConstant voltage, constant current, timer, pauseUser-selectable automatic power-up after powerfailure0–40 C10–95%, in the absence of condensation25 cm (L) x 21 cm (W) x 8 cm (H)1.1 kgFuse specifications are applicable to products with a serial number of 041BR300000 and higher.13

Appendix BWarranty and Ordering InformationWarrantyThe PowerPac Basic power supply is covered by a standard Bio-Rad Laboratorieswarranty. Contact your local Bio-Rad representative for details of the warranty. If anydefects should occur during this warranty period, Bio-Rad Laboratories will replace thedefective parts without charge. However, the following defects are specifically excluded:1. Defects caused by improper operation.2. Repair or modification done by anyone other than Bio-Rad Laboratories or theirauthorized agent.3. Use with cables or connectors not specified by Bio-Rad Laboratories for this powersupply.4. Deliberate or accidental misuse.5. Damage caused by disaster.For inquiry or request for repair service, contact your local Bio-Rad office.Warranty InformationModel:Serial Number:Date of Delivery:Warranty Period:Ordering ac Basic power supply, 100–240 VAC165-5061PowerPac Adaptor, qty 1165-5066PowerPac Adaptor, qty 212007433Replacement Fuse, 2.5 A, 250 V, 5 x 20 mm, fast-blow, qty 13Fuse specifications are applicable to products with a serial number of 041BR300000 and higher.15

Bio-RadLaboratories, Inc.Life ScienceGroup10000101114 Ver B (12009426)Web site bio-rad.com USA 1 800 424 6723 Australia 61 2 9914 2800 Austria 43 01 877 89019 Belgium 32 03 710 53 00Brazil 55 11 3065 7550 Canada 1 905 364 3435 China 86 21 6169 8500 Czech Republic 36 01 459 6192 Denmark 45 04 452 10 00Finland 35 08 980 422 00 France 33 01 479 593 00 Germany 49 089 3188 4393 Hong Kong 852 2789 3300 Hungary 36 01 459 6190India 91 124 4029300 Israel 972 03 963 6050 Italy 39 02 49486600 Japan 81 3 6361 7000 Korea 82 2 3473 4460 Mexico 52 555 488 7670The Netherlands 31 0 318 540 666 New Zealand 64 9 415 2280 Norway 47 0 233 841 30 Poland 36 01 459 6191 Portugal 351 21 4727717Russia 7 495 721 14 04 Singapore 65 6415 3188 South Africa 36 01 459 6193 Spain 34 091 49 06 580 Sweden 46 08 555 127 00Switzerland 41 0617 17 9555 Taiwan 886 2 2578 7189 Thailand 66 2 651 8311 United Arab Emirates 971 4 8187300United Kingdom 44 01923 47 1301US/EG18-09551218Sig 0118

PowerPac power supplies have passed test for operation at temperatures between 0º and 40º C, with relative humidity between 10 and 95% non-condensing. Operating the power supply outside these conditions is not recommended by Bio-Rad and will void the warranty. 1. To ensure adequate cooling of the power supply, be sure that there is at least 6 cmFile Size: 2MBPage Count: 26Explore furtherHow To Fix Bio Rad Power Supply Error Codes Tutorialloadware.orgPowerPac Basic Power Supply - Bio-Rad Laboratorieswww.bio-rad.comHow To Fix Bio Rad Power Supply Error 2 (Solved)loadware.orgPowerPac Basic Power Supply - Bio-Rad Laboratorieswww.bio-rad.comNO LOAD DETECTED---HELP! Scientist Solutionsscientistsolutions.comRecommended to you b