Endo-Leak TesterModel # ZUTR30005OPERATION MANUALZutron Medical, LLC9816 Pflumm RoadLenexa, Kansas 66215Phone 877-343-5873 Fax 913-967-5944ZUT-Lab-004-30005REV. D

Endo-Leak TesterModel # ZUTR30005Table of ContentsPage 2Introduction1. Intended Use2. Labels, Symbols and Signal Words3. Environmental Conditions4. Supply RatingsPage 3Important Safety InformationPage 4Mounting ConfigurationsCleaning ProcedurePage 5Getting to know the Endo-Leak TesterPage 6Getting to know the Endo-Leak Tester – Part IIPage 7Instructions For UseZutron Medical, LLC9816 Pflumm RoadLenexa, Kansas 66215Phone 877-343-5873 Fax 913-967-5944

Endo-Leak TesterModel # ZUTR30005IntroductionIntended UseThe Endo-Leak Tester is used as an early detection tool in wet leak testing to detect small holes in the exteriorof an endoscope. Early detection of leaks can prevent issues with cross contamination and costly repairs do tofluid invasion. Do not use this product for any other purpose.Compatible with ACMI, Fujifilm, Fuse, Olympus, Pentax, Richard Wolf & Storz endoscopes.Labels, Symbols and Signal WordsThe meanings of the labels, symbols and signal words that appear on the packaging, manual and/or instrumentare as follows:The following signal words are used throughout this manual:WARNINGIndicates a potentially hazardous situation which, if not avoided, could resultin death or serious injury.CAUTIONIndicates a potentially hazardous situation which, if not avoided, may resultin minor or moderate injury. It may also be used to alert against unsafepractices or potential equipment damage.NOTEIndicates additional helpful information.Environmental Conditions For indoor use onlyTemperature range: 10 C/50 F to 40 C/104 FSupply RatingsVoltage:DC 12V, 3.3ANOTEUse only the AC/DC power supply included with the unit.Zutron Medical, LLC9816 Pflumm RoadLenexa, Kansas 66215Phone 877-343-5873 Fax 913-967-5944Page 2

Endo-Leak TesterModel # ZUTR30005Important Safety InformationWARNINGPlug power cord into a GFI protected receptacle.WARNINGDo not immerse the unit in water.WARNINGDo not operate the unit if the cord is damaged or has been exposed towater.WARNINGInstall in accordance with electrical codes.WARNINGDo not open the air pump unit (electrical hazard).CAUTIONRead this entire manual before operating unit.CAUTIONAlways leak test an endoscope in accordance with manufacturer'sinstructions.CAUTIONAlways wear protective personal equipment when handling a dirtyendoscope.CAUTIONEnsure that the air hose connector is securely attached to the air pumpunit and that the endoscope leak test connector is fastened securely tothe endoscope. Ensure that the unit has been running for at least30 seconds before placing endoscope in water.CAUTIONDo not power the toggle switch off or press “STOP” while the endoscopeis submerged in water.CAUTIONAlways wait 10 seconds after powering toggle switch off to release the airfrom inside the endoscope before disconnecting the leak test connector.Zutron Medical, LLC9816 Pflumm RoadLenexa, Kansas 66215Phone 877-343-5873 Fax 913-967-5944Page 3

Endo-Leak TesterModel # ZUTR30005Mounting ConfigurationsStandard / Counter-TopThe standard Endo-Leak Tester has rubber bumpers on the bottom of the unit and is intended to be placed on aflat surface.Wall-MountThe Endo-Leak Tester has the option to be converted to a wall mount unit. A bracket that is sold separatelybracket may be installed by a trained service personnel. Please contact Zutron Medical for more information.CAUTIONPermanent wall-mounted installation should be performed by trainedservice personnel. Improper mounting could result in injury or damageto the Endo-Leak Tester.NOTEConsult with responsible personnel for proper location and bracing of themounting plate in your specific facility.Cleaning Procedure1. Lightly moisten a lint-free cloth with a working solution of a low foaming, neutral pH, medical gradedetergent that has been prepared in accordance with the detergent manufacturer’s instructions for use.2. Use the moistened cloth to wipe the exterior surfaces of the Endo-Leak Tester.3. Allow the detergent solution to remain on the exterior surfaces of the Endo-Leak Tester for the timerecommended by the detergent manufacturer.4. Lightly moisten a lint-free cloth with clean water and remove residual detergent from the surfaces of theEndo-Leak Tester.5. Thoroughly dry the surfaces of the Endo-Leak Tester with a dry lint-free cloth.6. Lightly moisten a lint free cloth with 70% ethyl or isopropyl alcohol.7. Use the moistened cloth to wipe the exterior surfaces of the Endo-Leak Tester.8. Allow the exterior surfaces of the Endo-Leak Tester to air dry.Zutron Medical, LLC9816 Pflumm RoadLenexa, Kansas 66215Phone 877-343-5873 Fax 913-967-5944Page 4

Endo-Leak TesterModel # ZUTR30005Getting to know the Endo-Leak TesterDigital DisplayStart / StopPower SupplyAir Hose ConnectionOn / Off SwitchAir HoseACMI Adapter - ZUTR10005AFujifilm Adapter - ZUTR10005FEndoscope AdapterOlympus / Fuse Adapter - ZUTR10005OPentax Adapter - ZUTR10005PRichard Wolf Adapter - ZUTR10005WStorz Adapter - ZUTR10005SZutron Medical, LLC9816 Pflumm RoadLenexa, Kansas 66215Phone 877-343-5873 Fax 913-967-5944Page 5

Endo-Leak TesterModel # ZUTR30005Getting to know the Endo-Leak Tester - Part IILeak Tester Adapter(See previous pagefor ordering information)Calibration GaugeZUTR10005HH(Sold separately or withEndo-Leak Tester KitCalibration Gauge AdapterZUTR10005CA(Sold separately or withEndo-Leak Tester KitDigital DisplayPower IndicatorLeak Tester Status(Stopped / Running)Set Output PressureOutput PressureMeasurementOutput PressureWhen RunningStart / Stop(Enter / Select In Settings)Menu / ServiceModeScroll Down / DecreaseSettingsScroll Up / IncreaseSettingsZutron Medical, LLC9816 Pflumm RoadLenexa, Kansas 66215Phone 877-343-5873 Fax 913-967-5944Page 6

Endo-Leak TesterModel # ZUTR30005Instructions For Use1.2.Plug in the power supply to the unit and attach the air hoseconnector to the front of the unit.Power the toggle switch on and press “RUN” to ensure nofluid is inside the tubing or endoscope adapter.After a few seconds press “STOP”.3.Before placing endoscope in water, attach the leak testadapter to the endoscope and press “RUN”.Wait at least 30 seconds before placing endoscope in water.4.Submerge endoscope in water and look for air bubbles thatindicate a leak coming from the endoscope, while deflectingthe bending section.5.After testing, remove the endoscope from water and press“STOP” on the Endo-Leak Tester unit.6.Always wait 10 seconds after powering the toggle switch offto release the air from inside the endoscope beforedisconnecting the leak test connector.Zutron Medical, LLC9816 Pflumm RoadLenexa, Kansas 66215Phone 877-343-5873 Fax 913-967-5944Page 7

5. Thoroughly dry the surfaces of the Endo-Leak Tester with a dry lint-free cloth. 6. Lightly moisten a lint free cloth with 70% ethyl or isopropyl alcohol. 7. Use the moistened cloth to wipe the exterior surfaces of the Endo-Leak Tester. 8. Allow the exterior surfaces of the Endo-Leak Tester to air